Automatic 0.1 Accuracy Electric Energy Meter Calibration Of Single Phase Energy Meter
This is a no-Ioad on-site test bench for single phase energy meters integrated with reference meter and source.
Main features
1. Accuracy class:0.1
2. Simultaneous measurement for 1-3 static single phase energy meters
3. Measurement for single phase local prepayment smart energy meters and single phase remote prepayment smart energy meters
4. Automatic/manual measurement of creeping and starting (of energy meters )
5. Wide measuring range reference meter integrated
6. Proofreading of constant by measuring pulse number,time,and start-stop power value;dial measurement according to number of KWh,time and number of pulse;preset number of KWh,the dial will stop accordingly
7. Automatic measurement of register
8. GPS reference clock integrated,time pairing for energy meters
9. Individual voltage loop for each testing position
Technical data
General |
|
Power supply |
220V±30% |
Power consumption |
<300VA |
Ambient temperature |
-20℃~40℃ |
Relative humidity (not condensing) |
40%~95% |
Dimension(L×W×H, mm) |
400×320×160 |
Weight |
<10Kg |
Warm-up time |
10min |
MTBF (Mean Time Between Failures) |
≥50000hrs |
Life time of whole set |
≥10Yrs |
LCD color display |
8”, 800×600 TFT |
Communication port |
RS232, USB |
Output voltage |
|
Measuring range |
30V~240V |
Adjusting range |
30V~240V continue adjustment |
Adjusting fineness |
0.002% |
Distortion |
≤2% |
Capacity |
≥10VA/position |
Output current |
|
Measuring range |
0.01~80A |
Adjusting range |
0.01~80A continue adjustment |
Adjusting fineness |
0.002% |
Distortion |
≤2% |
Capacity |
≥70VA/position |
Starting output current |
10mA |
Output power |
|
Stability |
≤0.03%/2min |
Output phase |
|
Adjusting range |
0°~360° |
Adjusting fineness |
0.01° |
Output frequency |
|
Adjusting range |
45~65Hz |
Adjusting fineness |
0.01Hz |
Reference meter |
|
Accuracy class |
0.1(direct connection), 0.3 (connected via clamp-on CT’s) |
Testing positions |
1~3 |
Monitoring |
|
Voltage |
Accuracy: 0.1 6 digits displayed Resolution: 0.1% |
Current |
Accuracy: 0.1 6 digits displayed Resolution: 0.1% |
Phase |
Accuracy: 0.2° 6 digits displayed Resolution: 0.2° |
Frequency |
Accuracy: 0.01Hz 6 digits displayed Resolution: 0.01Hz |
Standard deviation estimates |
≤0.02% |
Phase differences |
Voltage: ≤0.3 Current: ≤0.3 |
Reference time standard |
Frequency: 1Hz, 500kHz (TTL level) Stability: 2×10-8/day Accuracy: 2×10-7 |
Insulated resistant |
Output voltage/current loops to enclosure:>10MW Output voltage loop to current loop:>10MW Current to enclosure:>10MW |
Withstand voltage |
Output voltage/current loops to enclosure: 2000V/min, 50Hz Output voltage loop to current loop: 2000V/min, 50Hz Current to enclosure: 2000V/min, 50Hz |